Publisert 2021

Les på engelsk

Publikasjonsdetaljer

Utgiver : Nofima AS

Internasjonale standardnummer :
Trykt : 978-82-8296-699-3

Publikasjonstype : Nofimas rapportserie

Bidragsytere : Krakeli, Tor-Arne

Serier : Nofima rapportserie 36/2021

År : 2021

Forskningsområder

Kvalitet og målemetoder

Har du spørsmål om noe vedrørende publikasjonen, kan du kontakte Nofimas bibliotekleder.

Kjetil Aune
Bibliotekleder
kjetil.aune@nofima.no

Sammendrag

In the analysis ‘A 95 Molweight Distribution of Peptides’ BioLab use the Size Exclusion Chromatography (SEC) column Superdex Peptide. After a request for a new column of the same type it was informed that the column was no longer produced. The supplier recommended a replacement column, Superdex Increase. After a validation where the two columns was compared, it was concluded that BioLab could use Superdex Increase as a replacement for Superdex Peptide. This lead to a slight increase in the uncertainty for the methode, and a restriction in the measuring area. From ‘<200 – 20 000<’ to ‘<200 – 15 000<’.

Kontaktpersoner: